Magnetic Resonance Imaging and Radiation Exposure in Patients with Cardiovascular Implantable Electronic Devices

Publication Date: May 11, 2017

Key Points

Key Points

  • MRI generates static and gradient magnetic fields as well as RF energy. The potential interactions between CIEDs and electromagnetic interference from MRI include the following:
    • Magnetic field-induced force and torque due to ferromagnetic materials: CIED generator movement is extremely unlikely due to confinement in the subcutaneous tissues.
      • Leads do not contain any significant ferromagnetic materials to cause movement in a magnetic field.
    • Gradient magnetic field-induced electrical current: Gradient magnetic fields can induce current in conductive wires within the field that could lead to myocardial capture and potentially lead to atrial or ventricular arrhythmias.
    • Heating and tissue damage: RF fields can lead to nonconditional CIED component heating and subsequent thermal damage to the surrounding tissue (functional ablation). Changes in sensing or capture thresholds can occur as a result of tissue damage near lead electrodes.
    • Effects on reed switch activity: The reed switch is a feature that permits programming of the device by placement of a magnet. Magnetic fields might therefore affect the reed switch activity of a nonconditional CIED, leading to asynchronous pacing and inhibition of tachycardia therapies.
    • Electrical reset: High-energy electromagnetic interference (EMI) can lead to electrical or power-on-reset, a backup demand mode, wherein pacing might be inhibited and tachyarrhythmia therapy activated.
      • Power-on-reset parameters vary by vendor and type of CIED (See Tables 1 and 2), and can include reset of pacing polarity to unipolar.
      • Inhibition of pacing function due to oversensing of MRI-generated signals or pacing at an output below threshold (bipolar or unipolar) in a pacemaker (PM)-dependent patient might occur in the setting of power-on-reset and must be recognized to prevent catastrophic consequences.
      • Additionally, battery status can be affected, particularly for CIEDs that are near an elective replacement interval (ERI), which could result in unreliable function.
    • Inappropriate function and therapies: EMI from RF energy pulses or rapidly changing magnetic field gradients might cause oversensing that can lead to inappropriate inhibition of demand pacing and possibly asystole in a pacing-dependent patient, or induction of therapies such as inappropriate shocks in a patient with an implantable cardioverter defibrillator (ICD). Other inappropriate tracking or programming changes can occur.
  • These effects are influenced by various factors, including magnet field strength, RF power, position of the patient and the CIED within the MRI bore, CIED characteristics, and the size of the patient.

Table 1. Programmed Parameters for PMs During Power-On Reset Mode

Having trouble viewing table?

Manufacturer

Pacing mode

Pacing output

Pacing polarity

Sensitivity

Magnet1 response

BIOTRONIK

VVI 70 bpm

4.8 V @ 1.0 ms

Unipolar

2.5 mV

Yes

Boston Scientific2

VVI 65 bpm

5.0 V @1.0 ms

Bipolar

1.5 mV

No

Medtronic

VVI 65 bpm

5.0 V @ 0.4 ms

Bipolar

2.8 mV

Yes

St. Jude Medical

VVI 67.5 bpm

4.0 V @ 0.6 ms3

Unipolar

2.0 mV

No

ELA-Sorin

VVI 70 bpm

5.0 V @ 0.5 ms

Unipolar

2.2 mV

No

1 Magnet = device will/will not pace asynchronously in response to a magnet during safety mode/reset mode.
2 Boston Scientific CRT-P devices differ in pacing output (5 V @ 0.5 ms) and pacing polarity (right ventricle lead is unipolar and left ventricle lead paces from left ventricle to pulse generator).
3 St. Jude Medical Accent/Anthem and Frontier II models deliver 5 V @ 0.6 ms.

Table 2. Programmed Parameters for ICDs During Power-On Reset Mode

Having trouble viewing table?

Manufacturer

Rate cutoff

Detection criteria

Sensitivity

Energy

Pacing mode

Pacing output

BIOTRONIK

150 bpm

8/12

0.8 mV

40 J × 8

VVI 70 bpm

7.5 V @ 1.5 ms1

Boston Scientific

165 bpm

8/10

0.25 mV

41 J × 5

VVI 72.5 bpm

5.0 V @ 1.0 ms

Medtronic

188 bpm

18/24

0.3 mV

35 J × 6

VVI 65 bpm

6.0 V @ 1.5 ms

St. Jude Medical2

146 bpm

12

0.3 mV

36 J × 63

VVI 60 bpm

5.0 V @ 0.5 ms

ELA-Sorin

190 bpm

6/8

0.4 mV

42 J × 44

VVI 60 bpm

5.0 V @ 0.35 ms

All devices will respond to magnet application by temporarily disabling tachyarrhythmia detection. Pacing polarity for all devices is bipolar with the exception of Boston Scientific, which paces in a unipolar configuration. Energy values listed for Medtronic and St. Jude Medical represent energy delivered. The remaining represent energy charged.

1 In CRT devices, left ventricle lead output is 4.8 V @ 0.5 ms.
2 The St. Jude Medical Current and Promote family of devices revert to an autosense sensitivity setting, pace at VVI 67.5 beats per minute (bpm) with pacing outputs of 5.0 V @ 0.6 ms.
3 The St. Jude Medical Epic and Epic II family of devices delivers 30 J × 6.
4 ELA-Sorin LivaNova Ovatio family of devices: 34 J × 4.

Device Parameters

...vice Parameters...

...3. U.S. Food and Drug Administrat...


...nitions of CIED Systems in Relation to MR...


...ecommendations and Protocol for the Mana...

...al devices should be considered MR conditi...

...in a patient with an MR conditional system sh...

...ended for patients with an MR conditional syste...

...d for patients with an MR conditional...

...itative efforts and emergency treatment...

...ded for patients with an MR conditional system...

...le to perform an MR scan on a patient with an MR...


Management of Patients with a CIED Referred for MRI

...Patients with a CIED Referred for MRI...

...endations for the Decision to Perform...

...ble for patients with an MR noncondit...

...s reasonable to perform an MR scan i...

...ts with an MR nonconditional CIED, it i...


.... Recommendations for the Management of Pa...

...ommended for the patient with an MR...

...brillator/monitor (with external paci...

...ded that continuous MR conditional ECG and pul...

...commended that personnel with the skill t...

...or patients with an MR nonconditional CIED w...

...patients with an MR nonconditional CIED who are n...

...d that for the patient with an MR no...

...rrhythmia detections for patients with an...

...sible physician who is accountable for ove...

...s recommended that ECG and pulse ox...

...e efforts and emergency treatments that inv...

...nt with an MR nonconditional CIED wh...

...is reasonable to program patients with an...


...e 1. Management Algorit...


...e 2. Checklist for MRI Safety in the S...


...mplantable Loop Recorde...

...recommended that prior to MRI scanning...

...R scanning of MR conditional ILRs should be pe...


...Employee Safety...

...mended that the MR suite has a clearly delineated...


...10. Recommendations for the Management of Patients...

...s recommended that patients with a CIED...

...able to exclude the device from the field of view...

...might be reasonable to monitor patient...


Management of Patients with a CIED Undergoing CT Imaging

...of Patients with a CIED Undergoing CT...

...11. Secondary Neutron-Producing Rad...


...e 12. Recommendations and Protocol for the Managem...

...tiation of radiation treatment, a com...

...n-producing treatment is preferred over neutron-pr...

...weekly complete CIED evaluations for...

...D evaluation should be performed at the...

...visual and voice contact is recommended during e...

...on is recommended if its current location w...

...t be reasonable to perform a complete CI...

...IED relocation is NOT recommended...


Management of Patients with a CIED Undergoing Radiation Therapy

...ment of Patients with a CIED Undergoing Radiat...

...ist for Performance of Radiation Treatment...


...ED Management for Radiation Therapy...